Side Panel
ST80i Stress Test System Instructions for Use 2-13
Max (220)
Shows the maximum heart rate for patient, male or female.
BP mmhg
When the patient's NIBP is collected during the exercise stress test based on the defined
interval in the Exercise Protocol (Off, Begin, End, Every), the BP mmgh field displays the
measurement. If you click the
BP button during the patient session, the current blood pressure
measurement also displays in this field.
With this interface, there are visual and audio prompts (if enabled) for NIBP acquisition.
To enable audio alerts, see the ST80i Installation and Configuration Guide.
This field is also used to manually enter a blood pressure measurement, if an NIBP monitor is
not being used.
To override NIBP or manually insert the blood pressure measurement
NOTE Once the NIBP or BP is manually overwritten, ST80i no longer updates the real-time value from the
NIBP monitor.
1
Click the dashed lines or current BP value in the BP mmhg field.
2 Type in the blood pressure measurement.
3 Click OK.
Previous BP mmhg
When the blood pressure monitor records the current blood pressure measurement, the
Previous BP mmgh field displays the previous measurement. It will also display the previous
blood pressure measurement when you click the
BP button on the Procedure bar.
If you are using a blood pressure cuff, when you manually enter the blood pressure
measurement, this field records the previous blood pressure measurement.
SpO
2
The patient’s SpO
2
is collected in real time through a sensor. During the stress test, SpO
2
is
measured automatically and is continuously displayed on the screen. With this interface, there
are also visual and audio prompts (if enabled) for SpO
2
acquisition.
To enable audio alerts, see the ST80i Installation and Configuration Guide.
You can override the NIBP and SpO
2
values. Depending on the NIBP monitor used, you may
need to manually enter the SpO
2
data.