Table E-8 Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
Emissions
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment Guidance
Conducted
Immunity,
Induced RF
IEC
61000-4-6
3Vrms,
0.15 MHz to
80 MHz, @
10 Hz AM
3Vrms,
0.15 MHz to
80 MHz, @
10 Hz AM
Portable and mobile RF communications equipment
should be used no closer to any part of the ST80i
Stress Test System and optional Thermal Printer,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
Where
P is the maximum output power rating of the
transmitter in watts (
W) according to the transmitter
manufacturer and
d is the recommended separation
distance in meters (
m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
, should
be less than the compliance level in each frequency
range
b
.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Radiated RF
IEC
61000-4-3
General Test:
3V/m, 80 MHz
to 2.5 GHz @
10 Hz AM
Digital Radio
Test: 3V/m
80% AM @
1kHz ,
800 MHz to
960 MHz ,
1.4 GHz to
2.5 GHz
General Test:
3V/m, 80 MHz
to 2.5 GHz @
10 Hz AM
Digital Radio
Test: 3V/m
80% AM @
1kHz , 800 MHz
to 960 MHz ,
1.4 GHz to
2.5 GHz
P
Vrms
d
=
3
5.3
P
mV
d
=
/3
5.3
80 MHz to 800 MHz
P
mV
d
=
/3
7
800 MHz to 2.5 GHz