An Overview of the ST80i Stress Test System
2-26 ST80i Stress Test System Instructions for Use
When you record an event, ST80i generates a 12-lead ECG and documents the event name on
the ECG. The system also stores the event in memory and will print it in the Rhythm Events
portion of the final report. User Notes are also associated with the Event, if created. See
“Note” on page 2-28 regarding the Note icon.
In addition, ST80i automatically detects an arrhythmia event–if this is enabled. To enable
Arrhythmias, see the ST80i Installation and Configuration Guide. See also “Notifications and
Alerts” on page 3-36.
The default events include:
Supine
Mason-Likar
Standing
Hyperventilation
Chest Pain
Short of breath
To record an event
1
Click the Event button to display the Event drop-down list’
Figure 2-4 Event Drop-Down List
2
Select the event to record from the drop-down list.
The system prints an ECG and records the event for the final report.
To add an event
1
Click the Event button
2 Click Add New Event from the drop-down list to display the Add New Event pop-up box.