LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP Connector Replacement
Section Menu Section Contents Procedures Back Index
307
8
Removing the NIBP Connector
Refer to Figure 8.100 on p. 306.
To remove the NIBP connector from the parameter bezel (rear case—refer to Inside Rear Case Diagrams (p. 228)):
NOTE: These steps include all parameter bezel options. Your device may not have some of these options.
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Remove the system/therapy PCB assembly as described in System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232).
3. Remove the parameter bezel as described in Removing the Parameter Bezel (p. 311).
4. Remove the NIBP tube (3012180-02) from the NIBP connector (3207033-001).
5. Remove the fitting nut on the back side of the bezel to remove the connector (3207033-001).
6. Remove the NIBP connector from the bezel.