EasyManua.ls Logo

Physio Control LIFEPAK 15 - Page 325

Physio Control LIFEPAK 15
527 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Power/Contact PCB Cable (W05) Replacement
Section Menu Section Contents Procedures Back Index
327
8
Installing the Power/Contact PCB Cable (W05)
To install the power/contact PCB cable:
1. Connect the power/contact PCB cable (3207261-001) to J42 of contact PCB (A07).
2. Install the NIBP/CO2 assembly as described in Installing the NIBP (A21)/CO2 (A23) Modules (p. 292).
3. Install the OEM PCB as described in Installing the OEM PCB (A06) (p. 265).
4. Install the transfer relay assembly as described in Installing the Transfer Relay Assembly (A13) (p. 270).
5. Connect the power/contact PCB cable to the power PCB (A03) at J12 (may be labeled P12).
6. Install the system/therapy PCB assembly as described in Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239).
7. Reassemble the case as described in Reassembling the Case (p. 184).

Table of Contents

Other manuals for Physio Control LIFEPAK 15

Related product manuals