ROTALIGN
®
Ultra iS Shaft handbook
144
4. Adjust the laser beam as described in section 4.11.
5. Press
m then proceed to take measurement as described in section 4.12.
6. Press
e to stop measurement.
If the measurement mode used is not IntelliSWEEP or Continuous Sweep,
measurement is stopped by pressing
q and then highlighting ‘Stop’, and
confirming selection by pressing
e.
7. Switch both sensALIGN laser and sensor off then dismount them from the
presently measured coupling and mount them across the next coupling.
8. Press and hold
h/i to move to the cursor to the next coupling to be
measured. Observe the machine train inset at the left top corner of the display.
Please make sure when moving the laser and sensor to each coupling that the
dimension from sensor to coupling center is entered correctly in the dimensions
screen.
Always make sure the coupling you are measuring is the one actually highlighted
in the machine train!
9. Repeat the measurement procedure (steps 3 to 7) until measurements have been
taken across all couplings.
While in the measurement
screen, you may jump to
the desired coupling to be
measured by pressing and
holding
h/i.
Note
I
f
the measurement mode used is not IntelliSWEEP or Continuous Swee
p
,
measurement is stoppe
d
by
pressin
g
q
an
d
t
h
en
h
ig
hl
ig
h
ting ‘Stop’, an
d
con
rmin
g
selection b
y
pressin
g
e
.
Note
P
l
ease ma
k
e sure w
h
en movin
g
t
h
e
l
aser an
d
sensor to eac
h
coup
l
in
g
t
h
at t
h
e
dimension from sensor to couplin
g
center is entered correctl
y
in the dimensions
screen
.
A
l
wa
y
s ma
k
e sure t
h
e coup
l
in
g
y
ou are measurin
g
is t
h
e one actua
lly
h
i
ghl
i
gh
te
d
in
t
h
e
m
ac
hin
e
t
r
a
in!

Table of Contents

Related product manuals