117
5.2.11 Measurements options screen
The “Measurement options” screen may be accessed via the “Measurement” screen
context menu item ‘Measurement options. The context menu opens by pressing
q
while in the “Measurement” screen. Use
f/g and highlight ‘Measurement
options’.
With ‘Measurement options’ highlighted, press
e to confirm selection.
The “Measurement options” screen may be used to set the following selections:
› Whether Sweep measurement mode is started manually [by pressing
e]
or automatically when shafts are rotated
› Whether measurement points are taken manually [by pressing
e] or
automatically in Multipoint measurement mode – automatic collection of
measurement points is based on stability filter parameters and may be used in
Multipoint, IntelliPOINT and cardan shaft measurements
› Whether results resolution is set to two or three decimal places
› Whether the standard deviation is displayed on the screen or not
Alternatively, access the
“Measurement options”
screen directly using
the global menu. Press
q twice to open
the global menu, then
select ‘Measurement’ ->
‘Measurement options’.
Measurement options screen