57
After the scanning process is completed, any Bluetooth devices detected will be listed
on the screen.
Use
f/g to select the appropriate sensALIGN sensor (e.g.‘sensALIGN
49000134’). Press
e to confirm selection and proceed with measurement.
When sensALIGN sensor is initialized, and the laser beam is not centered, use the laser
adjustment pop-up wizard to correctly adjust the laser beam.
It is recommended to adjust the laser beam until the ‘Laser centred’ message appears
in the laser beam adjustment pane.
sensALIGN sensors that
have been detected are
stored in the system and
always appear in the sensor
selection screen. Sensors
may be deleted from the
displayed list by pressing
p with the sensor
highlighted.
The laser adjustment
pop-up wizard is the
primary laser beam
adjustment feature in the
Shaft application. Other
available features include:
1. Visual display of any
angular error in the
incident beam
2. XY view
3. Intelli laser position
4. Half function
These features are
described in section 5.2
‘Measurement options’.
Horizontal shaft alignment