cobas b 123 POC system 6 Measurement
Performing a measurement
q
Proficiency test materials, also known as EQA materials, have to be measured in the
Proficiency test menu.
u Performing a proficiency test measurement (p. 194)
You can measure samples from syringes (without cannulae), capillaries and the
Roche MICROSAMPLER PROTECT.
p To perform a syringe measurement
1
Collect and prepare your sample for the measurement.
Make sure that there are no air bubbles in your sample. Gently roll the syringe in
between your hands for 10-15 seconds.
Figure 6-8 Proper rolling technique for sample mixing
2
In the Overview tab, select the parameters that you want to measure.
Parameters colored in dark green will be measured.
Parameters colored in light green will not be measured.
Figure 6-9 Overview tab