162 Instructions for Use · Version 13.0
6 Measurement cobas b 123 POC system
Performing a measurement
3
To start the measurement, press the Start syringe measurement.
WARNING
Improper docking of the syringe or injecting a sample into the fill port during a
syringe measurement may result in infection
Injecting your sample into the fill port or improper docking of the syringe while
performing a syringe measurement may lead to the sample leaking from the fill port.
This may result in the transmission of blood and urine borne pathogens. In addition, it
may lead to a loss of sample.
r Do not inject the sample into the fill port while performing a syringe measurement.
r Do not hold onto the syringe after attaching it to the fill port.
r If a sample of human origin is spilled, wipe it up immediately and apply a
disinfectant.
r If a sample of human origin comes into contact with your skin, wash the affected
area immediately with soap and water, and apply a disinfectant. Consult a
physician.
r Firmly attach the syringe to the fill port.
4
Firmly attach the syringe to the fill port. Then, remove your hand from the
syringe.
Figure 6-10 Attaching the syringe to the fill port