cobas b 123 POC system 7 Quality control
Performing a proficiency test measurement
p To perform a proficiency test measurement
1
Workplace > QC measurement > Proficiency test
The following window appears on the screen:
Figure 7-18 Proficiency material setup screen
2
To enter the material name, level and lot number of the proficiency test material,
scan the barcode on the ampoule’s label.
Alternatively, you can manually enter this information by pressing
the button and using the keyboard box that appears on the screen.
3
Press the Start measurement button.
4
Make sure that the proficiency test material information that appears on the
screen is correct.
Figure 7-19 Proficiency test overview screen
5
To start the proficiency test, press the Start capillary measurement button.