7 Quality control cobas b 123 POC system
Performing a proficiency test measurement
6
Follow the instructions of the proficiency material manufacturer for handling and
preparation.
WARNING
Failure to follow proficiency test protocols may lead to incorrect proficiency
test results
Failure to follow proficiency test protocols may lead to incorrect proficiency test results.
r Follow proficiency test practices according to local regulations.
r Perform proficiency tests within 30 seconds of opening the proficiency test material
ampoule. Immediately discard the ampoule after performing the proficiency test.
r Use an ampoule adapter to carry out the proficiency test.
7
Firmly attach the ampoule adapter to the fill port.
Figure 7-20 Attaching the ampoule adapater to the fill port
8
To aspirate the contents of the ampoule, press the Ye s button.
Figure 7-21 Aspirating the contents from the ampoule