Roche Diagnostics
Instructions for Use · Version 13.0 197
cobas b 123 POC system 7 Quality control
Performing a proficiency test measurement
9
When you receive the prompt on the screen to remove the capillary, remove the
ampoule adapter from the fill port and press Ye s .
Figure 7-22 Removing the ampoule adapter from the fill port
The instrument performs the proficiency test.
Figure 7-23 Proficiency test values tab during a proficiency test
10
To enter the operator ID, press the button on the Operator ID field.