274 Instructions for Use · Version 13.0
10 Software functions cobas b 123 POC system
Configuring network connection settings
p To configure POCT1-A protocol settings
1
Utilities > Configuration > System > Connectivity > POCT1-A
Figure 10-33 POCT1-A configuration screen
2
To activate the POCT1-A protocol on your instrument, press
the and buttons to display Activated in the upper panel of the screen.
The bottom panel of the screen becomes active.
Figure 10-34 POCT1-A configuration screen
q
To send encrypted data using the POCT1-A protocol, activate the SSLv3/TLSv1 security
feature.
3
To activate SSLv3/TLSv1 security feature to the POCT1-A protocol, press
the and buttons on the Security field to display SSLv3/TLSv1.
q
The Keepalive feature automatically checks if there is an active connection between
your instrument and the corresponding information system (LIS or HIS).
4
To activate the Keepalive feature to the POCT1-A protocol, press
the and buttons on the Keepalive field to display Activated.
5
To enter an IP address for the POCT1-A protocol, press the button on the
IP address field. Enter your IP address using the numeric keyboard box that
appears on the screen. Then, press the OK button.