cobas b 123 POC system 10 Software functions
Configuring network connection settings
2
To activate the ASTM protocol on your instrument, press the and buttons
to display Activated in the upper panel of the screen.
The bottom panel of the screen becomes active.
Figure 10-32 ASTM configuration screen
q
To send encrypted data using the ASTM protocol, activate the SSLv3/TLSv1 security
feature.
3
To activate SSLv3/TLSv1 security feature to the ASTM protocol, press
the and buttons on the Security field to display SSLv3/TLSv1.
4
To enter an IP address for the ASTM protocol, press the button on the
IP address field. Enter your IP address using the numeric keyboard box that
appears on the screen. Then, press the OK button.
5
To enter a port number for the ASTM protocol, press the button on the Port
field. Enter your port number using the numeric keyboard box that appears on
the screen. Then, press the OK button.
6
To deactivate the ASTM protocol on your instrument, press
the and buttons to display Deactivated in the upper panel of the screen.
s