cobas b 123 POC system 10 Software functions
Configuring network connection settings
2
To enable remote operator management, make sure that the
Remote operator management button is checked and highlighted. If the
Remote operator management button is not checked and highlighted, press it.
s
p To add or remove parameters from an ASTM or a POCT1-A data
transmission
1
Utilities > Configuration > System > Connectivity > Parameter
Figure 10-38 Configuring ASTM and POCT1-A data transmissions screen
2
To select between the ASTM and the POCT1-A protocols, press
the and buttons on the upper part of the Select values for parameters
panel.
3
To add a parameter, choose it from the Available values list and press
the button.
4
To remove a parameter, choose it from the Selected values list and press
the button.
5
To add all parameters from the Available values list, press the button.
6
To remove all parameters from the Selected values list, press the button.
7
To copy your settings for one of the protocols and add it to the other protocol,
press the Copy to button.