306 Instructions for Use · Version 13.0
11 Consumable change cobas b 123 POC system
Changing consumables
q
A new AutoQC Pack has to be adjusted to room temperature at least 24 hours before use.
1
Workplace > Change Auto QC Pack
2
Follow the instructions on the screen.
q
Before inserting a new AutoQC Pack, disinfect the consumable area.
u To disinfect the instrument’s consumable area (p. 302)
s
p To change the printer paper
q
The printer paper is heat-sensitive on one side only. Ensure that you insert the paper roll
correctly, with the heat-sensitive side up.
1
Workplace > Change printer paper
2
Follow the instructions on the screen.