EasyManua.ls Logo

Rohde & Schwarz SMAB-B92 - Page 578

Rohde & Schwarz SMAB-B92
778 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMA100B
576User Manual 1178.3834.02 ─ 05
The COM/ID tone is sent according to the selected code, see Chapter D, "Morse Code
Settings", on page 740.
If no coding is set, the COM/ID tone is sent uncoded (key down).
Parameters:
<Code> string
Example: See Example "Configure and generate a VOR signal"
on page 544.
Manual operation: See "Code" on page 122
[:SOURce<hw>]:VOR:COMid:DASH
<Dash>
Sets the length of a Morse code dash.
Parameters:
<Dash> float
Range: 0.05 to 1
Increment: 1E-4
*RST: 0.3
Example: See Example "Configure and generate a VOR signal"
on page 544.
Manual operation: See "Dash Length" on page 123
[:SOURce<hw>]:VOR:COMid:DEPTh <Depth>
Sets the AM modulation depth of the COM/ID signal.
Parameters:
<Depth> float
Range: 0 to 100
Increment: 0.1
*RST: 10
Example: See Example "Configure and generate a VOR signal"
on page 544.
Manual operation: See "Depth" on page 123
[:SOURce<hw>]:VOR:COMid:DOT <Dot>
Sets the length of a Morse code dot.
If the time schema is set to standard, the dash length (= 3 times dot length), symbol
space (= dot length) and letter space (= 3 times dot length) is also determined by this
entry.
SOURce Subsystem

Table of Contents

Related product manuals