Remote Control Commands
R&S
®
SMA100B
578User Manual 1178.3834.02 ─ 05
Example: See Example "Configure and generate a VOR signal"
on page 544.
Manual operation: See "Period" on page 123
[:SOURce<hw>]:VOR:COMid:SYMBol <Symbol>
Sets the length of the Morse code symbol space.
Parameters:
<Symbol> float
Range: 0.05 to 1
Increment: 1E-4
*RST: 0.1
Example: See Example "Configure and generate a VOR signal"
on page 544.
Manual operation: See " Symbol Space " on page 124
[:SOURce<hw>]:VOR:COMid:TSCHema <Tschema>
Sets the time schema of the Morse code for the COM/ID signal.
Parameters:
<Tschema> STD | USER
STD
Activates the standard time schema of the Morse code.
The set dot length determines the dash length, which is 3
times the dot length.
USER
Activates the user-defined time schema of the Morse code.
Dot and dash length, as well as symbol and letter space
can be set separately.
*RST: STD
Example: See Example "Configure and generate a VOR signal"
on page 544.
Manual operation: See "Time Schema" on page 123
[:SOURce<hw>]:VOR:COMid[:STATe] <State>
Enables/disables the COM/ID signal.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Configure and generate a VOR signal"
on page 544.
SOURce Subsystem