Remote Control Commands
R&S
®
SMA100B
592User Manual 1178.3834.02 ─ 05
SWEep
Sets sweep mode.
The instrument processes frequency (and level) settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:FREQuency:STARt on page 595 and [:
SOURce<hw>]:FREQuency:STOP on page 595,
[:SOURce<hw>]:FREQuency:CENTer on page 594,
[:SOURce<hw>]:FREQuency:SPAN on page 595,
[:SOURce<hw>]:FREQuency:MANual on page 593
LIST
Sets list mode.
The instrument processes frequency and level settings by
means of values loaded from a list.
To configure list mode settings, use the commands of the Chap-
ter 13.16.7, "SOURce:LIST Subsystem", on page 612.
*RST:
CW
Example: See Example "Frequency configuration" on page 590
Example: See Example "Setup an RF frequency or power sweep"
on page 643
Manual operation: See "State (RF frequency sweep)" on page 171
[:SOURce<hw>]:FREQuency[:CW|FIXed] <Fixed>
Sets the frequency of the RF output signal in the selected path.
The effect depends on the selected mode:
In CW mode (FREQ:MODE CW | FIXed), the instrument operates at a fixed fre-
quency.
In sweep mode (FREQ:MODE SWE), the value applies to the sweep frequency.
The instrument processes the frequency settings in defined sweep steps.
In user mode (FREQ:STEP:MODE USER), you can vary the current frequency step
by step.
Parameters:
<Fixed> float
The following settings influence the value range:
An offset set with the command [:SOURce<hw>]:FREQuency:
OFFSet
Numerical value
Sets the frequency in CW and sweep mode
UP|DOWN
Varies the frequency step by step in user mode.
The frequency is increased or decreased by the value set with
the command [:SOURce<hw>]:FREQuency:STEP[:
INCRement].