Remote Control Commands
R&S
®
SMA100B
596User Manual 1178.3834.02 ─ 05
Parameters:
<Stop> float
Range: 300kHz to RFmax
Increment: 0.01Hz
*RST: 500 MHz
Default unit: Hz
Example: See Example "Setup an RF frequency or power sweep"
on page 643
Manual operation: See " Start Frequency/Stop Frequency " on page 176
[:SOURce<hw>]:FREQuency:STEP:MODE <Mode>
Defines the type of step size to vary the RF frequency at discrete steps with the com-
mands FREQ UP or FREQ DOWN.
Parameters:
<Mode> DECimal | USER
DECimal
Increases or decreases the level in steps of ten.
USER
Increases or decreases the level in increments, set with the
command FREQ:STEP[:INCR].
*RST: DECimal
Example:
// increasing the RF frequency with a step size of 50 KHz
SOURce1:FREQuency:STEP 50E3
SOURce1:FREQuency:STEP:MODE USER
SOURce1:FREQuency:CW UP
Manual operation: See "Variation Active" on page 66
[:SOURce<hw>]:FREQuency:STEP[:INCRement] <Increment>
Sets the step width.
You can use this value to vary the RF frequency with command FREQ UP or FREQ
DOWN, if you have activated FREQ:STEP:MODE USER.
Note: This value also applies to the step width of the rotary knob on the instrument
and, in user-defined step mode, increases or decreases the frequency.
Parameters:
<Increment> float
Range: 0 Hz to RFmax - 100 kHz
Increment: 0.01 Hz
*RST: 1E6
Example: See Example "Frequency configuration" on page 590
Manual operation: See "Variation Step" on page 66
SOURce Subsystem