EasyManua.ls Logo

Rohde & Schwarz SMAB-B92 - Page 610

Rohde & Schwarz SMAB-B92
778 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMA100B
608User Manual 1178.3834.02 ─ 05
Parameters:
<Period> float
Range: 1E-6 to 100
Increment: 1E-8
*RST: 1E-3
Example: See Example "Configuring the LF generator" on page 600.
Manual operation: See "Period" on page 101
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:RISE <Rise>
Selects the rise time for the trapezoid shape of the LF generator.
Parameters:
<Rise> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 250E-6
Example: See Example "Configuring the LF generator" on page 600.
Manual operation: See "Trapezoid Rise / Fall" on page 101
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:PERiod <Period>
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.
Parameters:
<Period> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 0.001
Example: See Example "Configuring the LF generator" on page 600.
Manual operation: See "Period" on page 101
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:RISE <Rise>
Selects the rise time for the triangle single of the LF generator.
Parameters:
<Rise> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 0.5E-3
Example: See Example "Configuring the LF generator" on page 600.
Manual operation: See "Triangle Rise" on page 101