Remote Control Commands
R&S
®
SMA100B
607User Manual 1178.3834.02 ─ 05
Manual operation: See "Period" on page 101
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:WIDTh <Width>
Sets the pulse width of the generated pulse.
Parameters:
<Width> float
Range: 1E-6 to 100
Increment: 1E-8
*RST: 5E-4
Example: See Example "Configuring the LF generator" on page 600.
Manual operation: See "Pulse Width" on page 101
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:FALL <Fall>
Selects the fall time for the trapezoid shape of the LF generator.
Parameters:
<Fall> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 250E-6
Example: See Example "Configuring the LF generator" on page 600.
Manual operation: See "Trapezoid Rise / Fall" on page 101
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:HIGH <High>
Sets the high time for the trapezoid signal of the LF generator.
Parameters:
<High> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 250E-6
Example: See Example "Configuring the LF generator" on page 600.
Manual operation: See "Trapezoid High" on page 101
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:PERiod <Period>
Sets the period of the generated trapezoid shape. The period determines the repetition
frequency of the internal signal.