Function Variants Crane Scale
6.4.5 Parameters
Table of parameters for the 'Crane Scale Function Variant'
Parameter Name Default Required
Setting
Recommended
Setting
Comments
Block 'Single Printout'
Timeout 0.0
0 sec: Print-out without no-motion state
monitoring
1-999 sec: Timeout to no-motion state
Printing option OFF
'OFF': No data output to the data
processing system and no
balancing
'to data
processing
system':
Data output to the data
processing system after
print-out
'Balance': Balance before print-out
'data processing
system/balance':
Balance before print-out
and data output to the data
processing system after
print-out
Block 'Print Total'
Timeout
Printing option
0.0
OFF
see above
Block 'Print Balance'
Timeout
Printing option
0.0
OFF
see above
Block 'Slack Rope'
FastComp.High
FastComp.Low
HIGH >
LOW
Slack rope detection.
Contact drops if weight falls below set OFF value.
The difference between ON/OFF produces hysteresis.
Block 'Overload'
FastComp.High
FastComp.Low
LOW >
HIGH
Overload protection device;
Contact drops if weight exceeds set OFF value.
The difference between ON/OFF produces hysteresis.
Block 'Double-Print Lock'
Type
ON
OFF
Percent
4%
5%
The "Crane Relieved" criterion is used to avoid
unintended double printout with double balancing.
Block 'High Lifting Speed'
Type
ON
OFF
Percent
110%
109%
If changeover
is required,
enter
corresponding
values.
Changeover load to be lifted; practically cut-off by
default setting.
BV-H2313 GB 0630 DISOMAT
®
Opus – Operating Manual
- 112 - © Schenck Process GmbH, Darmstadt