Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.3Reprocessing with the DAC UNIVERSAL ADVANCED
båÖäáëÜ
3. Open the snap lock on the basket and take the small lid off the basket.
4. Put maximum 5 solid instruments into the basket, if possible with the
sharp tip down. The solid instruments must be placed in the basket
so that the wire lid of the basket can be closed.
5. Attach the Class 5/6 chemical indicator to the metal bracket of the
BASKET lid. See "Using a chemical indicator with the BASKET lid
[ → 62]" for more information on using a chemical indicator.