5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
6. Place the small lid back on the basket.
7. Close the basket.
8. Insert the BASKET lid in the cover holder of the DAC UNIVERSAL in
such a way that the basket points down.
The solid instruments are in the BASKET lid; the DAC
UNIVERSAL is still open and ready for a cycle.
IMPORTANT
There is a metal cylinder in the middle of the small lid. The long part must
point toward you when closing.
IMPORTANT
Make sure the small lid of the basket closes easily and no instruments
are caught.
WARNING
Place a NitraClean tablet in the empty chamber. See "NitraClean
cleaning tablet [ 241]" for more information on using a NitraClean
tablet.
NOTICE
Make sure that the lid is inserted all the way into the holder so that the
basket does not touch the wall of the chamber. Make sure there are no
objects (other than the NitraClean tablet) in the chamber.
NOTICE
The DAC UNIVERSAL, the BASKET lid, and/or the instruments can be
damaged if the BASKET lid is not inserted correctly.