Sirona Dental A/S 7Display messages DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 7.1Display messages and what they mean DAC UNIVERSAL ADVANCED
båÖäáëÜ
74 When heating for
sterilization, a
temperature of 120°C /
248°F was not reached
within 2.5 minutes.
Increase the level of the "Ambient temperature
sensitivity" settings.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
If the DAC UNIVERSAL is in a room with a
temperature < 16°C / 61°F, ensure that operating
conditions is fullfilled.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC UNIVERSAL.
If the error occurs repeatedly, please contact your
authorized dealer.
75 Heating for sterilization
from 120°C / 248°F to
134°C / 273°F exceeded
2.5 minutes.
Increase the level of the "Ambient temperature
sensitivity" settings.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
If the DAC UNIVERSAL is in a room with a
temperature < 16°C / 61°F, ensure that operating
conditions is fullfilled.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC UNIVERSAL.
If the error occurs repeatedly, please contact your
authorized dealer.
76 Valve error Please contact your authorized dealer.
77 Valve error Please contact your authorized dealer.
Code
no.
Pictogram Description Corrective action