7Display messages DAC UNIVERSAL ADVANCED Sirona Dental A/S
7.1Display messages and what they mean DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
78 Valve error Please contact your authorized dealer.
79 Valve error Please contact your authorized dealer.
80 After reducing the water,
the desired sterilization
temperature was not
reached.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
Please contact your authorized dealer.
81 Low water level in the
water tank or oil bottle,
see DAC UNIVERSAL
display
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
Replace the NitramOil #2 bottle if it is empty, see
"Adding/replacing the lubrication concentrate".
If the error occurs repeatedly, please contact your
authorized dealer.
82 Error in the steam
generator
Please contact your authorized dealer.
83 The temperature sensor
in the chamber is
defective.
Please contact your authorized dealer.
Code
no.
Pictogram Description Corrective action