Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 8.1Regular activities
båÖäáëÜ
The appropriate indicator holder is used with a Class 5 chemical indicator,
when using the standard or sensitive program.
The appropriate indicator holder is used with a Class 6 chemical indicator,
when using the prion program.
A chemical indicator can be used in combination with the appropriate
indicator holder for batch control of the sterilization process of unwrapped
instruments.
For reprocessing of solid instruments no holder is needed, see „Using a
chemical indicator with the BASKET lid [ 156]“.
Chemical in dicator FLEX lid
While the FLEX lid is only for High-level thermal disinfection, the
instruments are exposed to the same sterilization process as performed
with the STANDARD lid. Therefore the same chemical indicators are
used for batch control of instruments reprocessed on the FLEX lid.
The manufacturer recommends using a chemical indicator for every
cycle.
The Class 5 chemical indicator monitors the time, temperature and
pressure parameters of the standard and sensitive programs.
A Class 6 chemical indicator monitors the time, temperature and pressure
parameters of the prion program.
8.1.2.1.1 Using a chemical indicator with the STANDARD lid
1. Attach the chemical indicator in the smaller of the two brackets on the
STANDARD indicator holder.
WARNING
If the chemical indicator indicates that the test was not passed, do not
use the instruments that were reprocessed after the time of the last
successful check.
Do not continue to use the DAC UNIVERSAL.
Please contact your authorized dealer.