60 53 362 D3472
156 D3472.201.10.20.02 07.2016
8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S
8.1Regular activities Operating Instructions DAC UNIVERSAL
2. Attach the STANDARD indicator holder in the groove on the center
post of the STANDARD lid.
3. Start a cycle. See "Reprocessing with the STANDARD lid DAC
UNIVERSAL STANDARD [ 172]" for the detailed procedure.
4. After the end of the cycle, check that the chemical indicator indicates
the sterilization test was passed.
8.1.2.1.2 Using a chemical indicator with the BASKET lid
1. Attach the chemical indicator to the metal bracket of the BASKET lid.
2. Start a cycle. See "Reprocessing with the BASKET lid DAC
UNIVERSAL STANDARD [ 177]" for the detailed procedure.
3. After the end of the cycle, check that the chemical indicator indicates
the sterilization test was passed.
8.1.2.1.3 Using a chemical indicator with the WRAPPED lid
1. Attach the chemical indicator in the smaller of the two brackets on the
STANDARD indicator holder.