Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD
båÖäáëÜ
17. Remove the BASKET lid from the holder on the DAC UNIVERSAL at
the end of the cycle to allow the instruments to cool down faster.
18. Place the BASKET lid on the counter or put it in the optional lid holder.
19. Check the instruments visually to make sure they are dry.
20. Check the instruments visually to make sure they have been cleaned.
21. Check the cycle number, properly conducted cycle on your
documentation (printer, software, etc.): For sterilization, "Sterilization,
time, ok" must be indicated.
22. Check whether the dark status bar of the chemical indicator is in the
"ACCEPT" range.
If everything is correct, reprocessing was successful.
23. Document the cycle according to the work instructions applicable in
your practice.
24. Open the snap lock on the basket and take the small lid off the basket.
25. Remove the solid instruments from the BASKET lid.
26. Place the small lid back on the basket.
The solid instruments can now be used again.
Sterilization of unwrapped instruments is not sufficient for invasive
procedures. After packing the instruments, see "Pre- and post-
reprocessing of instruments [ → 153]". another sterilization must be
conducted in the DAC UNIVERSAL or a class B sterilizer (e.g. DAC
PROFESSIONAL). For sterilizing wrapped instruments in DAC
UNIVERSAL, see "Reprocessing with the WRAPPED lid DAC
UNIVERSAL STANDARD [ → 182]".
After the end of the cycle, the DAC UNIVERSAL automatically
reactivates lubrication for the next cycle.
CAUTION
Danger of burning!
Even when the BASKET lid is half or completely open, the instruments
are still hot. Allow the extremely hot instruments to cool down after the
process.
CAUTION
Danger of burning!
The metal part of the lid is very hot. The plastic part of the basket can be
hot. Please touch only the plastic part of the BASKET lid at the end of
the cycle.