60 53 362 D3472
180 D3472.201.10.20.02 07.2016
8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S
8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
Depending on the program selected, the following message may
appear on the display:
"Ready 134 °C"
,
"Ready 121 °C"
or
"Ready 134 °C 18min"
.
12. Press the "Start"-Key.
A prompt to select the lid will appear on the display.
13. Select the BASKET lid using the arrows keys
14. Press the „Start“-Key twice.
On the DAC UNIVERSAL, only the LED at the left of the "Rinse/
Flush"-Key is on.
Depending on the program selected, a prompt to press down on
the safety ring may appear on the display.
15. Press down on the safety ring to start the cycle. If the safety ring is
not pressed within 20 seconds, you must choose the lid and press the
"Start"-Key again.
The BASKET lid of the DAC UNIVERSAL closes as soon as the
safety ring is pressed.
The DAC UNIVERSAL automatically runs the cleaning and
sterilization cycle.
You can see on the display at any time which phase the DAC
UNIVERSAL is in.
The following message appears on the display when the selected
cycle is finished:
"Press C to unload"
. The lid moves about
halfway out of the chamber. This indicates that a cycle has been
completed and the instruments have been sterilized.
16. Press the "Clear"-Key.
The BASKET lid will open completely.
IMPORTANT
If no program is selected, the standard program for reprocessing is
used.
CAUTION
Danger of burning!
Never hold your head directly over the DAC UNIVERSAL or touch the
inserted lid, while the DAC UNIVERSAL opens, as steam comes out of
the unit upon completion of a cycle.
WARNING
After completion of the cycle, make sure that all solid instruments are
still in the basket. If instruments have fallen into the chamber during the
cycle, cleaning and sterilization have failed. In this case, reprocessing of
the instruments must be repeated.