8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S
8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
5. After placing the instrument wrapped in a pouch in the wire brackets,
pull the instrument lightly to make sure that it is properly attached
between the brackets.
6. Insert the WRAPPED lid in the cover holder of the DAC UNIVERSAL.
The instruments are on the WRAPPED lid; the DAC UNIVERSAL
is still open and ready for a cycle.
7. Press the "Wrapped"-Key to select the cycle fro wrapped
instruments.
On the DAC UNIVERSAL, only the LED at the left of the
"Wrapped"-Key is on.
The display shows: Choose lid type WRAPPED
8. Press the "Start"-Key.
The display reads:
"Start wrapped cycle?Instr. treat in std.cycle
before wrap? Press "M" to confirm"
.
9. Press the "Menu/Enter"-Key to confirm the selection.
Depending on the program selected, a prompt to press down on
the safety ring may appear on the display.
10. Press down on the safety ring to start the cycle. If the safety ring is
not pressed within 20 seconds, you must choose the lid and press the
"Start"-Key again.
The WRAPPED lid of the DAC UNIVERSAL closes as soon as
the safety ring is pressed.
The DAC UNIVERSAL automatically runs the standard program
without cleaning or lubrication.
You can see on the display at any time which phase the DAC
UNIVERSAL is in.
The following message appears on the display when the selected
cycle is finished:
"Press C to unload"
. The lid moves about
halfway out of the chamber. This indicates that a cycle has been
completed and the instruments have been sterilized.
NOTICE
Make sure that the lid is inserted all the way into the holder and the
wrapped instruments do not touch the wall of the chamber. Make sure
there are no objects in the chamber.
NOTICE
The DAC UNIVERSAL and/or the instruments can be damaged if the
WRAPPED lid is not inserted correctly.
WARNING
If no program is selected, the standard program for reprocessing is
used.