Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD
båÖäáëÜ
11. Press the "Clear"-Key.
The WRAPPED lid will open completely.
12. Remove the WRAPPED lid from the holder on the DAC UNIVERSAL
at the end of the cycle to allow the instruments to cool down faster.
13. Place the WRAPPED lid on the counter or put it in the optional lid
holder.
14. Check the instruments visually to make sure they are dry.
15. Check the instruments visually to make sure they have been cleaned.
16. Check the cycle number, properly conducted cycle on your
documentation (printer, software, etc.): For sterilization, "Sterilization,
time, ok" must be indicated.
17. Check whether the dark status bar of the chemical indicator is in the
"ACCEPT" range.
If everything is correct, reprocessing was successful.
18. Document the cycle according to the work instructions applicable in
your practice.
19. Remove every instrument from the WRAPPED lid. Pull every
wrapped instrument singly from the respective bracket.
CAUTION
Danger of burning!
Never hold your head directly over the DAC UNIVERSAL or touch the
inserted lid, while the DAC UNIVERSAL opens, as steam comes out of
the unit upon completion of a cycle.
WARNING
After completion of the cycle, make sure that all instruments are still
properly attached to the WRAPPED lid. If instruments have fallen into
the chamber during the cycle, sterilization has failed. In this case, the
cycle must be repeated.
CAUTION
Danger of burning!
Even when the WRAPPED lid is half or completely open, the
instruments are still hot. Allow the extremely hot instruments to cool
down after the process.
CAUTION
Danger of burning!
The metal part of the lid is very hot. Please touch only the plastic part of
the WRAPPED lid at the end of the cycle.