Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED
båÖäáëÜ
Manual oil operation
Oil can be manually pumped to the chamber by selecting this option. This
can be used to ensure that the DAC UNIVERSAL is able to lubricate the
instruments correctly.
Please see section „Checking oil supply to the chamber of the DAC
UNIVERSAL ADVANCED [ → 122]“ for further instructions.
Manual water operation
Water can be manually pumped to the chamber by selecting this option.
This can be used to ensure that the DAC UNIVERSAL is able to bring
water to the chamber correctly.
Please see section „Checking water supply to the chamber of the DAC
UNIVERSAL ADVANCED“ for further instructions.
Manual water channels:
Water can be manually pumped through the bottom adapter in the
chamber by selecting this option. This can be used to ensure that the
DAC UNIVERSAL is able to bring water to the internal water channels of
the STANDARD and FLEX lids.
Please see section „Checking water supply to the lids of the DAC
UNIVERSAL ADVANCED [ → 123]“ for further instructions.
Drain tank
The water tank can be drained through the water inlet on the backside of
the autoclave. If there’s an automatic water system connected to the inlet,
this must be removed first.
IMPORTANT
After using the manual oil operation the chamber and connecting hoses
needs to be cleaned for oil residue, see section "Drain the chamber" in
"Drain the water tank via the menu [ → 125]".
IMPORTANT
After using the manual water operation the chamber and connecting
hoses needs to be cleaned for water residue, see section "Drain the
chamber" in "Drain the water tank via the menu [ → 125]".
IMPORTANT
After using the manual water operation the chamber and connecting
hoses needs to be cleaned for water residue, see section "Drain the
chamber" in "Drain the water tank via the menu [ → 125]".