5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
Drain chamber
Drains the chamber through the drain outlet. Select this function, if there’s
residue in the chamber that needs to be drained. An empty lid must be
inserted into the autoclave.
About 60 seconds after starting this operation the lid will reopen.
The chamber should then be empty. If not, the operation can be repeated.
Flow Test
Measures the supply air quantity.
5.2.2.3.3 Settings
The settings section of the menu is used to alter the user changeable
settings of the DAC UNIVERSAL. Some of these settings alter the way
the cycles are completed. Please always be careful when altering these
settings.
5.2.2.3.3.1 Oil settings
The oil settings make it possible to change the level of lubrication the
instruments receive during a cycle.
Level 1: standard setting
Level 2: 25% more oil than level 1
Level 3: 50% more oil than level 1