Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED
båÖäáëÜ
5.2.2.3.3.2 Setting ambient sensitivity
Standard setting: Level 1 (low ambient temperature sensitivity).
Level 0: ambient temperature sensitivity deactivated (For clinics with
a high / normal temperature).
Level 1: Low ambient temperature sensitivity.
Level 2: Medium ambient temperature sensitivity.
Level 3: High ambient temperature sensitivity.
Level 4: Maximum ambient temperature sensitivity (for clinics with a
very low temperature).
When there is a low temperature in the practice/clinic, we recommend
increasing the ambient temperature sensitivity somewhat – see
„Requirements of the location / Ambient conditions [ 49]".
Please note that activating or increasing the level of sensitivity can
potentially prolong the total cycle time and increase power consumption.