60 53 362 D3472
88 D3472.201.10.20.02 07.2016
5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
5.2.2.3.3.3 Printer settings
The printer settings enable the user to control the printer attached to the
serial port of the DAC UNIVERSAL.
These settings are:
● Printer ON:
If a printer is connected to the autoclave, the cycle info will be printed.
● Printer OFF:
Nothing will be printed
● Repeat last print:
The last print will be printed again.
● Printer language settings:
Through these options, the user is able to select which language the
print should be printed in. The available languages are English,
German & French.
5.2.2.3.3.4 Rinse cycles
The number of rinse cycles performed doing a total cycle can be adjusted,
using these settings.
Standard setting: 4 rinse cycles (3 cold + 1 warm).
This can be increased to a maximum of 6 rinse cycles (5 cold + 1 warm)
The "Flex" lid will always use 6 rinse cycles (5 cold + 1 warm).
Please note that the total cycle time is prolonged and water consumption
increased by using more than 4 rinse cycles.