Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED
båÖäáëÜ
5.2.2.3.3.5 Oil expel
Condensation and excess oil are purged from the instruments with air.
Standard setting: Level 1.
The maximum is level 5.
Please note that oil is purged before sterilization and that increasing oil
purging prolongs the total cycle time.
Level 1 is recommended for instruments of the manufacturers W&H,
KaVo, NSK, and most other instruments.
Level 3 is recommended for Sirona T1 CLASSIC/TE.
However, please observe the manufacturer's instructions for use with
each instrument.
5.2.2.3.3.6 Expel
Purge instruments with air after sterilization.
Standard setting: 0. The maximum is level 3.
● Level 0: no blowing
● Level 1: 50 sec. effective blowing time
● Level 2: 100 sec. effective blowing time
● Level 3: 150 sec. effective blowing time
Please note that increasing the blowing time prolongs the total cycle time.
Level 1 is recommended for instruments of the manufacturers W&H,
KaVo, NSK, and most other instruments.
Level 3 is recommended for Sirona T1 CLASSIC/TE.
However, please observe the manufacturer's instructions for use for each
instrument.
WARNING
If processing instruments for invasive surgical treatments, Expel
function cannot be used without prior installation of a sterile-filter.