Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED
båÖäáëÜ
5.2.2.3.3.5 Oil expel
Condensation and excess oil are purged from the instruments with air.
Standard setting: Level 1.
The maximum is level 5.
Please note that oil is purged before sterilization and that increasing oil
purging prolongs the total cycle time.
Level 1 is recommended for instruments of the manufacturers W&H,
KaVo, NSK, and most other instruments.
Level 3 is recommended for Sirona T1 CLASSIC/TE.
However, please observe the manufacturer's instructions for use with
each instrument.
5.2.2.3.3.6 Expel
Purge instruments with air after sterilization.
Standard setting: 0. The maximum is level 3.
Level 0: no blowing
Level 1: 50 sec. effective blowing time
Level 2: 100 sec. effective blowing time
Level 3: 150 sec. effective blowing time
Please note that increasing the blowing time prolongs the total cycle time.
Level 1 is recommended for instruments of the manufacturers W&H,
KaVo, NSK, and most other instruments.
Level 3 is recommended for Sirona T1 CLASSIC/TE.
However, please observe the manufacturer's instructions for use for each
instrument.
WARNING
If processing instruments for invasive surgical treatments, Expel
function cannot be used without prior installation of a sterile-filter.