Scale System Error Codes
Error Code Definition and Recommended Action
Error - E16 The analog-to-digital converter (ADC) response timed out or the measurement exceeds the capacity
of the ADC. This error may occur during normal scale system operation. If this error did not occur
during in conjunction with “Error - E03 in. a defective control board or load cell may the cause of the
error.
• Verify the load cell cables are properly connected to the control board.
• Verify the load cell cables are not cut or severed.
• Contact Technical Support if the problem persists.
Error - E17 The weight measurement taken exceeds 999.9 lb or kg. A defective control board or load cell may be
the cause of this error. This error may occur during normal scale system operation.
• Verify the load cell cables are properly connected to the control board.
• Verify the load cell cables are not cut or severed.
• Contact Technical Support if the problem persists.
Note
Location of the cells in relation to the number it is associated with in the cell verification module:
Cell 0 - Foot Right
Cell 1 - Head Right
Cell 2 - Head Left
Cell 3 - Foot Left