Screen that enables setting the reverberation time
measurement parameters: Start Delay, Method, Octave,
Freq. Range, Recording Time, Time Step, Averaging,
Noise Mar., Logger Name and Level.
Delay between pressing the <Start> key and the start of
measurement integration.
Method of the RT60 calculation: Decay (Interrupted
Noise Method) or Impulse (Impulse Response Method).
The selection of the method depends on the used type of
the sound source.
Method of RT60 calculation that uses the impulse sound
source (like pistol shot, petard explosion).
Method of RT60 calculation intended for measurements
when room is excited by broad or narrow band sound
noise source (usually pink noise).
Type of spectrum (1/1-octave or 1/3-octave) based on
which the RT60 analysis is performed.
Frequency range for RT60 calculations: 63Hz-4kHz (7
bands) and 63Hz-16kHz (9 bands) for 1/1-octave; 50Hz-
5kHz (21 bands) and 50Hz-20kHz (27 bands) for 1/3
octave.
Time of measurement data (sound pressure level decay
curve) registration during RT60 calculations: 1s .. 30 s.
Time-step of data registration (sound pressure level) in
the file during RT60 calculations: 2, 5, 10, 20, 50,
100 ms.