173
SVAN 977A User Manual - Appendixes
Word
number
Name
Comment
6
Range
measurement range:
1 LOW
2 HIGH
7
StatDet
SLM mode:
detector's type for statistical analysis:
0 Linear,
1 Exponential
VLM mode:
reserved
8
RepCycle
repetition cycle:
0 infinity
nnnn number of repetitions (1..1000)
9
NofProf
number of profiles: 3
10
StartDelay
start delay time specified in seconds:
0 ÷ 60
11..12
IntTimeSec
integration time specified in seconds
13
StartSync
Synchronization the start of measurement with RTC
0 switched off.
-1 synchronization to 1 sec.
1 synchronization to 1 min.
15 synchronization to 15 min.
30 synchronization to 30 min.
60 synchronization to 1 hour.
14
LeqInt
detector's type in the RMS/LEQ function:
0 Linear,
1 Exponential
15
SpectrumBuff
1/x OCTAVE logger (sum of the following flags):
8 LEQ/RMS spectrum
FFT logger (sum of the following flags):
8 averaged FFT spectrum
in other cases:
reserved
16
MainResBuff
Summary results. Contents defined as a sum of:
0 none
1 Main Results
2 Spectrum
4 Spectrum Max
8 Spectrum Min
16 Spectrum Peak
32 Statistical levels
64 Statistical analysis in profiles
128 Statistical analysis in 1/x OCTAVE mode
256 RPM
512 Meteo

Table of Contents