Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J A-15
When the possibility exists that multiple calibration methods are
available, one must be selected to achieve a span number. All other
calibration methods should be material factored to the proven span
number.
Calibration Method Availability
R-Cal Built-in
Static weights Optional item
Test chain Optional item
Other Material Factoringpre- or post-
weighed material. (User supplied.)
Since all Micro-Tech models are equipped with R-Cal, this example will
assume an R-Cal span #. It then will factor the static weights to the R-
Cal span number. Other factoring is similar and not shown.
To reset the existing weight material factor, do the following.
1. Press the Menu button twice to bring up “Main Menu 2.”
Material
Factoring
Reset Weight
Factor
BATCH
- MAIN MENU 2 -
Press MENU for more
READY
ALARM
CALIB
SCALE
DATA
DISPLAY
CALIB
DATA