B-60 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
This section lists the technical details for all parameters and functions in
the Micro-Tech’s “Main Menu 6” screen.
Press the AUDIT TRAIL soft key on Main Menu 6 to access the
function.
The Audit Trails function is a method for recording all changes in setup
and calibration data. Any time you change a value that affects weighing,
the new parameter is recorded with time and date of the change. All
changes can be printed when required.
To view Audit Trails press the Edit button and, using the keypad, enter
the trail number (from 1 to 999) that you want to view. Then press the
Enter button.
Here are the technical specifications for the Linearization menu. The
Linearization function is not normally used with conveyor-belt scales. In
case of extreme instances where calibration at multiple flow rates may
be required, please consult a Thermo Fisher Scientific technical
representative.
08-13-2013 4:18p
VAR. S1 Zero
O 32356 N 32396