B-32 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
If you do not enter the correct password, the following screen appears.
Pressing RETURN returns the system to the previous function. If you
enter the correct password, the previous screen appears and access is
allowed.
When the protection level is changed using the on line procedure, the
system automatically returns to protected status if no keyboard entries
are made within 60 seconds.
Here is the Protection menu table.
Parameter
Selection
Description
PROTECTION
LEVEL
NONE,
LIMITED,
PROTECTED
The PROT menu is activated when a
Service or Operator password is
entered.
Sets protection level.
Required to enter the password in order
to decrease the protection level.
NONE The system is completely
unprotected; all data can be read or
changed.
LIMITED Operator functions and data
are protected. All setup and calibration
data are protected except zero
calibrate.
PROTECTED The system is totally
protected; process data can be read,
but no changes are allowed.
Protection Menu
Table
BATCH
- SYSTEM PROTECTED -
INVALID PASSWORD
ACCESS DENIED
READY
ALARM
CALIB
RETURN