EasyManua.ls Logo

Thermo Scientific Micro-Tech 9101 - Installing the Panel Model

Thermo Scientific Micro-Tech 9101
329 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction
Hardware Installation
1-10 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
8. In the case of sourcing power for the AC outputs/inputs from the
integrator, source the power from the AUX PWR OUT terminal.
9. All additional field wiring operation at voltages less than 30 V must
be located on the left bottom of the enclosure. Leave ample loose
wiring (typically 8 inches / 20 cm) to facilitate removing the terminal
connectors.
10. Close and latch the enclosure door.
Figure 1–4. Connectors on Underside of Enclosure
This model of the Micro-Tech is designed to be mounted in an
instrument panel. The instrument panel should not be exposed to
excessive vibration, heat, or moisture. The front bezel, when properly
seated, forms a dust seal. A two-inch clearance around the top and
bottom of the Micro-Tech is required for convection cooling. Additional
clearances may be required if other equipment mounted directly below
the Micro-Tech generates excessive heat. A 2-3 inch (50-75mm)
clearance in the back is necessary for wiring access and fuse
replacement. A 1-inch (25mm) clearance on each side is necessary for
inserting the chassis-holding brackets from the back after inserting the
Micro-Tech.
Installing the
Panel Model

Table of Contents

Related product manuals