Operation
Calibrating the Micro-Tech
Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J 3-9
To perform an R-Cal span calibration for your scale, do the following.
1. Make sure the Run screen (see Figure 3–2) is currently being
displayed.
2. Press the Menu button and the “Main Menu 1” screen appears.
3. Press the Span Calibration button and the following screen appears.
If the R-Cal auto-span option (shown above) is not displayed, go to
page B-5 to change the span calibration method. (For additional
information about the Calibration menu, go to page B-22.)
Doing an R-Cal
Span Calibration
Press MENU for more
Run the belt empty, then
press START