Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J B-33
Here is the Diagnostic menu table.
1 A/D gross 1
A/D gross 2*
A/D gross 3**
A/D gross 4**
A/D net
0 - 131070 Digital representation of gross load at the
load cell.
*Available only when Type of scale set to
Two A/D Channel.
**Available only when Type of scale set to
Four A/D Channel.
Digital representation of “material” load at the
load cell. (gross – zero No.)
#1-J16____mV
#2-J21____mV
1
#3-J1_____mV
2
#4-J2_____mV
2
#J9_____mV
3
Gross load at load cell. Presented in mvdc
format.
Note: After A/D conversion.
1
Available only when Type of scale set to
Two A/D Channel.
2
Available only when Type of scale set to
Four A/D Channel.
3
Model 9201.
10000)
Fine adjust of load cell mv/v zero set. Factory
adjusted. Not to be changed in the field
Fine adjust of load cell mv/v span set.
Factory adjusted. Not to be changed in the
#1_____ pls/min
A divider to bring the speed pulses to within
a range of 1500-2500 pulses per minute.
Incoming speed pulses (after prescale) per
minute
pulses
Total number of speed pulses during a test
duration (calibration time). Not applicable for
simulated speed.
length
Test duration total belt length. Not applicable
for simulated speed.
password
Enter Service password to enable protection.
Up to eight characters long.
It is strongly recommended you write down
the password and keep a copy in a safe
place.
To remove a password, just press EDIT and
then ENTER. (Password: Service.)
password
Enter Operator password to enable
protection.
Up to eight characters long.
It is strongly recommended you write down