Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J A-19
2. Press the Start button and the following screen appears.
3. Make sure the belt is running empty, then press the Start button. The
following Auto Spanning screen appears.
4. When the Auto Span procedure is completed, the span percent-error
is displayed.
Record the results and note whether the error is positive or negative.
Time remaining 360
Rate xxx.x Tph
Tot xx.xx Tons
Run belt, then
press START
Change Span?
Error x.xx %