Initializing the Micro-Tech
Initializing the Software
Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J 2-19
NOTE. Ignore the “Memory Erased” message. The date and time you
already entered have been retained.
2. To select the current language, press the Enter button and the scale
set-up screen appears.
3. To choose another language, repeatedly press the Choice button until
the language you want is displayed, then press the Enter button.
This menu allows you to enter the number of load cells and A/D
(analog/digital) channels you are using with your Micro-Tech. The
default value for the Micro-Tech is determined by the belt-scale code.
1. The Micro-Tech display should currently look like this.
2. The Micro-Tech menus are also known as the Micro-Tech “scrolls.”
Please go to Figure 3–1 and note that the Micro-Tech keypad
contains an up-scroll button and a down-scroll button, which are also
known as the up-arrow and down-arrow buttons. Thus, the notation
in the display saying “Press down SCROLL,” is a cue to press the
down-arrow (or down-scroll) button, as described the next step
(step 3).
3. Press the down-arrow button (or Scroll button) and the “Scale Data
Scroll 1” screen appears.
Entering Scale
Data
BATCH
Initial scale setup
and calibration
Press down SCROLL
READY
ALARM
CALIB